PaX-i (PCH-2500) User Manual
87
7. Acquiring CEPH Images
NOTE
For information regarding the exposure time of each CEPH
imaging mode, refer to the section ‘3.4.2 CEPH mode’.
D. The gender and age of the patient
are automatically selected according to
EasyDent’s patient information. Nevertheless,
ensure that this information is correct.
Age Group VATECH’s Standard
Child
≤ 12
Adult
Man
≥ 13
Woman
NOTE
A Child is defined as a person who is younger than 12
years old. If Child is selected, the image size and exposure
dose are automatically reduced.
E. Select X-ray intensity.
NOTE
X-ray intensity (Hard, Normal, Soft) is according to the
operator's decision.
Soft ≤ Normal ≤ Hard
F. Default values for the tube voltage (kVp)
and current (mA) will be displayed based
on the patient’s
on the patient’s gender and X-ray intensity. If necessary, you may manually adjust
these values using the arrows to the right of each value.
NOTE
Adjustable resolution:
- Tube voltage: ± 1 kVp
- Tube current: ± 1 mA
G. Click
Conrm
to accept the parameters.
Pi_u272_en_20180514.indd 87 2018-05-14 오전 11:03:27

Table of Contents

Other manuals for Vatech PAX-I

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech PAX-I and is the answer not in the manual?

Vatech PAX-I Specifications

General IconGeneral
Focal Spot Size0.5 mm
Image ReceptorDigital sensor
CompatibilityDICOM 3.0
Image DetectorCMOS
Power Supply220-240 VAC, 50/60 Hz
Sensor TypeCMOS
X-ray GeneratorHigh Frequency
Tube Voltage60-90 kVp
Tube Current4-12 mA
Power Requirements100-240V AC, 50/60Hz
UsageDental diagnostics

Related product manuals