EasyManua.ls Logo

Yamatake STD910 - Figure 7-3 Connection for Calibration

Default Icon
256 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance and Troubleshooting Yamatake Corporation
7-14 ST3000 Smart Transmitter Series 900 Electronic Differential Pressure/Pressure Transmitter
Set up for calibration
Wire the transmitter in a similar way to that shown below.
Figure 7-3 Connection for calibration
Set range
Before starting calibration work, use the SFC to check that the set range of the trans-
mitter agrees with the specifications. If they do not correspond, use the SFC to set the
correct range. Procedures are found in "6-7-8 : Display or Change Low/High Limits
and Span of Set Range".
The set range is calibrated assuming that the low limit 20 kPa and the high limit is 100
kPa.
Black
Red
High-precision
ammeter
High-pressure side
Standard pressure
generator
Open the low-pressure
side to the air.
Power supply cable
24V DC
power supply
SFC

Table of Contents

Related product manuals