Yamatake Corporation Operation of the Transmitter
ST3000 Smart Transmitter Series 900 Electronic Differential Pressure/Pressure Transmitter 5-11
If the screen display is not 0 kPa in step 3 or 0% in step 4, proceed to step 5 to perform
zero-point calibration.
Assuming that 0.005 kPa was displayed in step 3, perform zero point calibration, using
this procedure:
4
Press the key.
Step Description SFC screen
5
Press the and keys, in
that order
6
Press the key.
7
Press the key.
Step Description SFC screen
OUT-
J
PUT
INPUT
OUTPUT FIT-1234
WORKING...
OUTPUT FIT-1234
0.00 %
SHIFT
OUT-
J
PUT
INPUT
SHIFT-
INPUT FIT-1234
WORKING...
INPUT FIT-1234
0.0000 kPa
COR-
K
RECT
RESET
INPUT FIT-1234
ZERO INPUT?
ENTER
( Yes )
NON-VOL
INPUT
ARE YOU SURE?