5-36 IM 760301-01E
Setting the Integration Timer
5. Press the Timer Setting soft key to display the Timer Setting dialog box.
6. Press the cursor keys to select the hour, minute, or second box.
7. Press SET to display the entry box.
8. Press the cursor keys to set the hour, minute, or second that you selected in
step 6 to 0.
9. Press SET or ESC to close the box.
10. Repeat steps 6 to 9 to set the hour, minute, and second to 00000:00:00.
Performing Integration in the Manual Integration Mode
Starting the Integration
11. Press INTEG to display the Integ menu.
12. Press the START soft key. The START indicator to the right of the INTEG key
illuminates, and the integration is started.
Holding the Integration
13. Press HOLD. The HOLD key illuminates, and the numeric data display is held.
Integration continues inside the WT3000.
Releasing the Hold Operation
14. Press HOLD while the values are held. The HOLD key turns OFF, and the
numeric data display is updated. In the hold condition, the display can be
updated by carrying out single measurements (by pressing SINGLE).
Stopping the Integration
15. Press the STOP soft key. The START indicator turns OFF and the STOP
indicator illuminates. The integration time and integration value are held.
Integration continues if you press START at this point.
Resetting the Integration
16. Press the RESET soft key. The STOP indicator turns OFF, the integration time
and integration value are reset, and the display of integration measurement
functions such as WP shows [-------] (no data).
5.12 Setting the Manual Integration

Table of Contents