OI/SCK
-EN Rev. E Operating instruction SCK Gas Sampling System 151
The following tables show the maintenance intervals within which to perform
maintenance work, with a link to the relevant instructions:
Maintenance task See … Inspection/Maintenance
intervall
as required
dayiy
weekly
twice yearly
yearly
every 2 yea
rs
General tasks
Check the system and clean
if necessary
Checking the condition of the system,
cleaning the system
on page 163
X
Display and evaluate trend
displays
Evaluating trend displays
on page 166
X
Functional test of residual
current circuit breaker
Checking the residual current circuit
breaker
on page 167
X
Functional tests of:
EMERGENCY STOP
switch
Warning lamps
Warning buzzer
external warning horn (op-
tional)
Checking protective installations
on
page 169
X
Maintenance task See … Inspection/Maintenance
intervall
as required
dayiy
weekly
twice yearly
yearly
every 2 years
Gas sampling probe H
Mechanical cleaning of
probe pipe and tip
Gas sampling probe H: Removing
incrustations
on page 176
X X
Replace intake filter
Gas sampling probe H: Replacing the
intake filter
on page 178
X
X
Replace membrane pump
membranes
Gas sampling probe H: Replacing
membrane pump membranes
on page
181
X
Maintenance plan

Table of Contents

Other manuals for ABB SCK