SensyMaster FMT430, FMT450 THERMAL MASS FLOWMETER | OI/FMT430/450-EN REV. B 109
Menu / parameter
Description
Device Setup / ...Transmitter
...Units
Selection of submenu ‘...Units
’ using .
...Custom Units
Selection of submenu ‘...Custom Units’ using .
TX Location TAG
Enter the measuring point tagging for the transmitter.
Alphanumeric, max. 20 characters
TX TAG
Entry of the TAG number for the transmitter.
Alphanumeric, max. 20 characters
Perform Device Reset
For service purposes only.
Restart the device without having to switch the power supply on and off.
Factory Defaults
All user-accessible parameters will be reset to the factory default settings.
...Feature Settings
Selection of submenu ‘...Feature Settings
’ using .
Device Setup / ...Transmitter / ...Units
Unit Massflow Qm
Selection of unit for mass flow.
Refer to
Table 2: Units for the mass flow on page 96.
The selection a
pplies to the display of the current mass flow, and for the parameters related to mass flow such as Qm
Max
and Qm
Max
DN.
Mass Totalizer
Select the unit for the mass totalizer.
Refer to Table 7: Units for the mass totalizer on page 96.
Unit Volumeflow Qv@
Selection of unit for the standard volume flow.
Refer
to Table 1: Units for the standard volume flow on page 96.
The selection applies to the display of the current standard volume flow and for the parameters related to standard volume
flow such as
Qv@ Max and Qv@ Max. DN.
Unit Vol.@ totalizer
Select t
he unit for the standard volume totalizer.
Refer to Table 8: Units for the standard volume totalizer on page 97.
Std. Conditions Vol@
Selection of the standard state for standard volume flow.
Refer to
Table 4: Standard conditions on page 96.
Special Cond. fact.
Enter the factor by which the standard state 0°C 1atm is multiplied to define an individual standard state.
Volumeflow Qv@ Name
Enter the name of the unit for the standard volume flow.
If the first character is a space, a @ sign will be appended to the unit (standard setting).
Alphanumeric, maximum 7
characters.
Volume@ Tot. Name
Enter the name of the unit for the standard volume totalizer.
If the first character is a space, a @ sign will be appended to the unit (standard setting).
Alphanumeric, maximum 7 characters.
Temperature
Selection of unit for
temperature.
Refer to
Table 5: Temperature units on page 96.
Pressure
Select a unit for pressure.
Refer to Table 9: Pressure units on page 97.
Length
Select a unit for length information.
Refer to
Table 6: Length units on page 96.
Density@
Select a unit for standard density.
Refer to Table 3: Standard density units on page 96.
Density@ Name
Enter the name of the unit for standard density.
If the first
character is a space, a @ sign will be appended to the unit (standard setting).
Alphanumeric, maximum 7
characters.