Alarms 55
fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22
#
Alarm Text
Alarm Type
Cause
Solution
Main
Priority
Sub
Priority
17.
Charge battery
(<15min)
Electrical power
Ventilator is not
connected to mains
supply.
Remaining battery
time < 15 min.
Switch to mains
supply.
The device
continues in battery
mode without
interruption.
Alarm can be
silenced with Alarm
Silence button.
I
2
18.
Flat battery!
Electrical power
Remaining battery time
< 1 min. Ventilation
can no longer continue
on battery power
Immediately switch to
Mains supply or
Access alternative
ventilation
I
2
19.
check Blender
System alarms
Proportional mixer or
control defective
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
2
20.
Patient
disconnected
Disconnection
Tubes blocked
Leak or disconnection.
Check tubing
system for leaks.
Check setting of
Pinsp.
I
3
21.
flow sensor
calibration
required
Sensor alarms
Flow sensor is not
calibrated
Calibrate flow sensor
I
4
22.
O
2
value out of
range
Sensor alarms
O
2
level is out of range
Calibrate O
2
sensor
and or disconnect the
device from AC power
and contact your local
Acutronic Medical
Systems Distributor
I
4
23.
Clean flow
sensor
Sensor alarms
Water or secretion in
flow sensor
Clean replace flow
sensor.
I
4
24.
Flow sensor
defect
Sensor alarms
Damaged heating
wires inside flow
sensor.
Replace flow sensor.
I
4
25.
Flow sensor
not connected
Sensor alarms
Flow sensor not
connected.
Defective sensor
cable.
Check flow sensor and
sensor cable
connection.
I
4
26.
O
2
sensor
calibration
failed
Sensor alarms
Error occurred during
calibration.
Repeat calibration.
I
4
27.
Oxygen sensor
defect
Sensor alarms
O
2
sensor defective.
Replace O
2
sensor.
I
4
28.
Oxygen sensor
used up
Sensor alarms
O
2
sensor worn.
Replace O
2
sensor as
quickly as possible.
Calibration still
available.
I
4