56 Alarms
fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22
#
Alarm Text
Alarm Type
Cause
Solution
Main
Priority
Sub
Priority
29.
O
2
value out of
range
Sensor alarms
O
2
sensor or mixer
defective.
Repeat Calibration
and/or Replace O
2
sensor.
Contact your local
ACUTRONIC
Medical Systems
Distributor.
I
4
30.
SpO
2
module
disconnected
Sensor alarms
SpO
2
module is not
connected
Connect SpO
2
module
I
5
31.
SpO
2
sensor
failure (XY)
Sensor alarms
XY internal reference:
- SpO
2
sensor is
faulty
- 38: SpO
2
module
DSP software
version not
supported
- 62: SpO
2
board
failure
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
5
32.
check SpO
2
sensor
Sensor alarms
SPO
2
sensor is off
patient or
malfunctioning
Connect SpO
2
sensor
to patient
I
5
33.
Apnea alarm
Patient alarms
No spontaneous
patient breathing
detected.
Check patient,
Switch to controlled
ventilation.
I
6
34.
low SIQ
Patient alarms
SPO
2
sensor has low
SIQe (Signal Quality)
Check SpO
2
sensor for
correct contact with
patient or replace
sensor
I
7
35.
FiO
2
at max
Patient alarms
FiO
2
has reached
maximum setting
Adjust settings
I
7
36.
Lower
minimum FiO
2
Patient alarms
Lower minimum FIO
2
than set for PRICO
Adjust settings
I
7
37.
High pressure
Patient alarms
Pressure rise in tubing
system; mechanical
inspiration was
reduced to relieve the
system.
Check patient,
Check tubing
system.
Replace patient
system.
I
7
38.
High PEEP
Patient alarms
PEEP 6 mbar above
set value for at least 15
seconds.
Check patient,
Check patient
connection,
Check tubing.
Adjust settings
I
7
39.
High SpO
2
Patient alarms
SpO
2
measured
above set limit
Check patient,
Check SpO
2
sensor
for correct contact
with patient
II
8

Table of Contents