Alarms 57
fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22
#
Alarm Text
Alarm Type
Cause
Solution
Main
Priority
Sub
Priority
Adjust settings
40.
Low SpO
2
Patient alarms
SpO
2
measured below
set limit
Check patient,
Check SpO
2
sensor
for correct contact
with patient
Adjust settings
II
8
41.
High Pulse
Rate
Patient alarms
Pulse rate measured
above set limit
Check patient,
Check SpO
2
sensor
for correct contact
with patient
Adjust settings
II
8
42.
Low Pulse
Rate
Patient alarms
Pulse rate measured
below set limit
Check patient,
Check SpO
2
sensor
for correct contact
with patient
Adjust settings
II
8
43.
Low Perfusion
Index
Patient alarms
Low SpO
2
perfusion
index
Check patient,
Check SpO
2
sensor
for correct contact
with patient
Adjust settings
II
8
44.
Low PIP
Patient alarms
Leak or disconnection.
Check patient,
Check patient
connection.
Check tubing
system for leaks.
Check setting of
Pinsp.
II
8
45.
Charge battery
(<60min)
Electrical power
Ventilator is not
connected to mains
supply.
Power failure.
Internal battery
remaining charge
less than 60
minutes.
Switch to mains
supply.
Alarm can be silenced
with Alarm Silence
button. The device
continues in battery
mode without
interruption.
II
8
46.
Charge battery
(<30min)
Electrical power
Ventilator is not
connected to mains
supply.
Remaining battery
time < 30 min.
Switch to mains
supply.
Alarm can be silenced
with Alarm Silence
button. The device
continues in battery
mode without
interruption.
II
8
47.
High breath
rate
Patient alarms
Hyperventilation
self-trigger
Check patient,
Adjust frequency
II
8