13. Hold the Movable Plate on which the
body of the Sample Probe Motor is
mounted. Remove the two set-screws
located on the sides of the Osmometer
Module Top Plate. The Movable Plate
will be free to drop under its own weight
upon removal of the set-screws. Gently,
lower the Movable Plate until it rests on
top of the Stripper Bracket (
Figure 8A &
Figure 8B
).
14. Remove the two socket head cap screws
and washers that mount the Sample Probe
Motor to the Movable Plate (
Figure 9).
15. Mount the replacement Sample Probe
Motor to the Movable Plate using the two
socket head cap screws and washers.
Figure 7
Figure 8A
Figure 9
112104RPM Rev0
(OsmoPRO Service Manual)
Page 3 of 5
Figure 6B
Two Set
Screws
Hex Nut
Three Cable
Clips
Figure 8B
Two
Socket
Head Cap
Screws