Connector into the Control PCB.
29. Secure the C10 Sample Probe Cable and
the C11 Block Probe Cable to the side of
the Power Supply & Controls Assembly
using a Cable Clamp and a screw.
30. Reverse steps one through twelve of the
General Disassembly Instructions to re-
assemble the instrument.
112104RPM Rev0
(OsmoPRO Service Manual)
Page 5 of 5