6. Disconnect the following Connectors on
the Cooling System Harness from their
mating Connectors: A1, A2, A3 & A4.
7. Rest the instrument on its back side.
Remove the two screws that secure the
Power Supply & Controls Assembly to
the bottom of the Chassis (
Figure 3).
8. Place the instrument in the upright posi-
tion. Remove the two screws that secure
the Power Supply & Controls Assembly
to the rear of the Chassis (
Figure 4).
9. Slide the Power Supply & Controls
Assembly out of the Base Assembly.
10. Remove the four socket head cap screws
and washers that mount the Osmometer
Module Mounting Bracket to the Chassis.
Carefully, slide the Osmometer Module
out of the instrument (
Figure 5A & Figure
5B
).
11. Remove the hex nut and washer that
secure the SPG & BPG Ring Terminals
to the side of the Osmometer Module
Mounting Bracket (
Figure 6).
12. Remove the four socket head cap screws
and washers that mount the Osmometer
Module to the Osmometer Module
Mounting Bracket, which is secured to
the Chassis. Lift the Osmometer Module
off of the Osmometer Module Mounting
Bracket. (
Figure 7A & Figure 7B).
Figure 3
Figure 4
Figure 6
Figure 5A
112406RPM Rev0
(OsmoPRO Service Manual)
Page 2 of 3
Two Screws
Two Screws
Figure 5B
Four
Socket
Head
Cap
Screws
Hex Nut