13. Mount the replacement Osmometer
Module to the Osmometer Module
Mounting Bracket using the four socket
head cap screws and washers.
14. Secure the SPG & BPG Ring Terminals
to the side of the Osmometer Module
Mounting Bracket using the hex nut and
washer.
15. Slide the Osmometer Module into the
instrument, and mount the Osmometer
Module Mounting Bracket to the Chassis
using the four socket head cap screws
and washers.
16. Slide the Power Supply & Controls
Assembly into the Base Assembly.
17. Secure the Power Supply & Controls
Assembly to the rear of the Chassis using
the two screws.
18. Rest the instrument on its back side.
Secure the Power Supply & Controls
Assembly to the bottom of the Chassis
using the two screws. Return the instru-
ment to the upright position.
19. Pair the following Connectors on the
Cooling System Harness with their mat-
ing Connectors: A1, A2, A3 & A4.
20. Attach the following Connectors on the
Motor/Sensor Cable Harness to their mat-
ing Connectors: B1, B2, B3, B4, B6, B7,
B8, B9 & B10.
21. Plug the C10 and C11 Connectors into
the Control PCB.
22. Affix the C3G Ring Terminal to the side
of the Power Supply & Controls
Assembly using a screw. Plug the C3
Connector into the Control PCB.
23. Secure the C10 Sample Probe Cable and
the C11 Block Probe Cable to the side of
the Power Supply & Controls Assembly
using a Cable Clamp and screw.
24. Reverse steps one through twelve of the
General Disassembly Instructions to re-
assemble the instrument.
Figure 7A
112406RPM Rev0
(OsmoPRO Service Manual)
Page 3 of 3