4
ALC
5
4. While monitoring the internal detector, the RF mo dulator level
is varied until the detected p ower is equivalent to the reference
power measured in step 2.
5. Mo dulation is re-enabled if appropriate.
These steps are performed in approximately 200

s and are repeated
any time p ower or frequency is changed.
See Also
Softkeys listed ab ove,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Fltness Menu
,
4
MOD
5
,
4
POWER LEVEL
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Set Atten
\Externally Leveling the Swept Signal Generator" in Chapter 1
\Working with Mixers/Reverse Power Eects" in Chapter 1
\Working with Sp ectrum Analyzers/Reverse Power Eects" in
Chapter 1
Operating and Programming Reference A-9