Doubler Amp Mode On
Function Group
POWER
Menu Map
5
Description
This softkey is applicable to instrument mo dels with a doubler
installed. The doubler has an integral amplier whose op eration is
controlled by the instrument rmware. This softkey turns o the
automatic mo de of op eration and turns on the amplier so that it is
always used. This is an
unspeciedmode
of operation since it can
cause increased harmonics and degraded dynamic range at some
frequencies. This softkey has no eect on instruments without a
doubler.
An asterisk next to the key lab el indicates that this feature is active.
Programming Codes
SCPI:
POWer:AMPLier:STATE ON
j
OFF
j
0
j
1
POWer:AMPLier:STATE?
Analyzer:
NONE
See Also
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Dblr Amp Menu
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
Dwell Coupled
Function Group
FREQUENCY
Menu Map
2
Description
This softkey lets you couple the dwell time for p oints in the stepped
frequency sweep mo de to the ramp sweep mo de sweep time. The
equation to determine the dwell time in the dwell coupled mo de is as
follows:
Coupled Dwell Time = (sweep time)
4
(number of step p oints)
An asterisk next to the key lab el indicates that this feature is active.
Programming Codes
SCPI:
SWEep[:FREQuency]:DWEL1:AUTO ON
j
1
Analyzer:
NONE
See Also
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Step Swp Menu
D-10 Operating and Programming Reference