This hardkey allows up to eight dierent front panel settings to b e
stored in memory registers 1 through 8. Swept signal generator
settings can then b e recalled with the
4
RECALL
5
key. A memory
register can b e alternated with the current front panel setting using
the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Atrnate Regs
softkey.
The information stored in memory registers is retained in memory
indenitely when ac line p ower is constantly available, or for
approximately three years without line p ower. Pressing
4
PRESET
5
does
not erase the memory registers (1 through 8).
Register 0 is a memory register also. It sa
ves the last front panel
settings automatically and can not b e accessed through the
4
SAVE
5
key. Likewise, register 9 is reserved for user preset storage and can
not b e accessed with the
4
SAVE
5
key. Pressing
4
PRESET
5
erases register
0, but not register 9.
Programming Codes
SCPI:
*SAV
<
num
>
The above is an IEEE 488.2 common command.
Analyzer:
SVn, where n= a numeric value from 1 to 8.
See Also
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Altrnate Regs
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Clear Memory
,
4
RECALL
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Save Lock
\Saving and Recalling an Instrument State" in Chapter 1
\Programming Typical Measurements" in Chapter 1
Operating and Programming Reference S-1