NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Freq Offset
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
Freq Offset
Function Group
FREQUENCY
Menu Map
2
Description
This softkey lets you set a frequency oset value and applies it to
all frequency parameters. The frequency oset ranges between and
including
6
110.0 GHz. Changing the frequency oset value changes
the display but does not aect the output frequency.Frequency
multiplier and oset are related as shown by the following equation:
Entered value or Displayed Frequency = (Frequency Generated
2
Multiplier) + Oset value
The factory preset value is 0 Hz. An asterisk next to the key lab el
indicates that this feature is active.
Programming Codes
SCPI:
FREQuency:OFFSet
<
num
>
j
MAXimum
j
MINimum
FREQuency:OFFSet:STATeON
j
OFF
j
1
j
0
Analyzer:
SHFB
<
n
>
[Hz
j
Kz
j
Mz
j
Gz
j
]
See Also
FREQUENCY
4
MENU
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Freq Mult
Operating and Programming Reference F-19