ALC BW Menu
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ALC Bandwidth Select Low
Sets the ALC bandwidth to the low
bandwidth p osition (10 kHz), and
to remain there for all sweep and
mo dulation conditions.
See Also
4
ALC
5
\Optimizing Swept Signal Generator Performance" in Chapter 1
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
Altrnate Regs
Function Group
SYSTEM
Menu Map
8
Description
This softkey causes the swept signal generator to alternate on
successivesweeps between the present instrument state and a second
instrument state stored in an internal register (1 to 8). Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Altrnate Regs
once to turn it on, a second time to turn it o. An
asterisk next to the key lab el indicates that this feature is active.
Programming Codes
SCPI:
SYSTem:ALTernate:STATeON
j
OFF
j
1
j
0
SYSTem:ALTernate
<
num
>
j
MAXimum
j
MINimum
Analyzer:
AL1n, where n= 1 through 8 function on, AL0 function o
See Also
4
RECALL
5
,
4
SAVE
5
\Saving and Recalling an Instrument State" in Chapter 1
A-12 Operating and Programming Reference