Auto Fill Incr
Function Group
FREQUENCY, POWER
Menu Map
2, 5
Description
This softkey is used in two lo cations:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Fltness Menu
and
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
List Menu
.
Flatness Menu - When selected, the swept signal generator waits for
a frequency incrementvalue to be entered.
--> Increment:
is
displayed in the active entry area. A list of frequencies is created
automatically, b eginning at the auto ll start frequency and always
ending with the auto ll stop frequency. The swept signal generator
uses the incrementvalue on all points, but if the stop frequency
requires a dierent increment to b e used to b e exact, the swept
signal generator simply ends the frequency list at the stop frequency
disregarding the incrementvalue.
If the incrementvalue requested creates a list that exceeds the
number of elements available, the follo wing message app ears:
TOO MANY CORRECTION PTS REQUESTED
List Menu - When selected, the swept signal generator waits for
a frequency incrementvalue to be entered.
--> Increment:
is
displayed in the active entry area. A list of frequencies is created
automatically, with all p oints separated by the frequency increment
value. The list b egins at the auto ll start frequency and ends at a
frequency less than or equal to the auto ll stop frequency
.
If the incrementvalue requested creates a list that exceeds the
number of points available (801), the following message appears:
TOO MANY LIST PTS REQUESTED
Programming Codes
SCPI:
NONE, see
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Fltness Menu
or
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
List Menu
Analyzer:
NONE
See Also
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Fltness Menu
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
List Menu
\Optimizing Swept Signal Generator Performance" in Chapter 1
A-24 Operating and Programming Reference