SCPI COMMAND SUMMARY
ABORt
Causes the sweep in progress to ab ort and reset. If
INIT:CONT
is
ON
it immediately restarts the sweep. The p ending op eration ag
(driving *OPC, *WAI, and *OPC?) undergoes a transition once the
sweep is reset.
AM[:DEPTh] <num>[PCT]|MAXimum|MINimum|<num>DB
AM[:DEPTh]? [MAXimum|MINimum]
Sets and queries the percentage of AM when the SOURce is
INTernal. If
<
num
>
is received with units of dB, the value is
converted to percentby the equation:
[%] = 100 * (1
0
10**(
0
dB/20))
Valid ranges of dB are 0 to 40 dB. After *RST, the value is 50%.
AM:INTernal:FREQuency <num>[<freqsuffix>]|MAXimum|MINimum
AM:INTernal:FREQuency? [MAXimum|MINimum]
Sets and queries the frequency (in Hz) of the internal AM source.
After *RST, the value is 1 kHz.
AM:INTernal:FUNCtion SINusoid|SQUare|TRIangle|RAMP|NOISe
AM:INTernal:FUNCtion?
Sets and queries the waveform of the internal AM source. After
*RST, the value is SINusoid.
AM:SOURce INTernal|EXTernal
AM:SOURce?
Sets and queries the source of the AM mo dulating signal. After
*RST, the value is EXTernal.
AM:MODE DEEP|NORMal
AM:MODE?
Controls the AM depth limits of the swept signal generator. The
NORMal
position is selected at *RST.
AM:STATe ON|OFF|1|0
AM:STATe?
Sets and queries the status of the AM mo dulation. After *RST, the
setting is
OFF
.
AM:TYPE LINear|EXPonential
AM:TYPE?
Sets and queries the type of AM mo dulation. After *RST, the
setting is
LINear
.
CALibration:AM:AUTO ON|OFF|1|0
CALibration:AM:AUTO?
Sets and queries the automatic mo dulator calibration switc
h.
Operating and Programming Reference S-27