OPTION NOT INSTALLED. The language selected and the
corresp onding rmware/hardware necessary to run that language
is not present in the swept signal generator. See Chapter 3 for
information on language selection.
DISPLAY IS NOT RESPONDING
: Can app ear on the front panel
emulator if the internal processor can not communicate with the
display prop erly. This error indicates a display failure or a display
connector problem.
DIVIDE BY ZERO EXCEPTION
: This can only b e caused by
an internal pro cessor error. Refer to Chapter 4 for instructions on
contacting a qualied service technician.
EEROM FAILED, LOST CAL
: This error indicates that the swept
signal generator has lost its calibration constants and may not meet
specications. Refer to Chapter 4 and follow the lo cal op erator's
check pro cedures. If you are a qualied service technician and this
failure o ccurs, read the Calibration Constants section in the
Agilent
Technologies 8360 B-Series Swept Signal Gener
ator/8360 L-Series
Swept CW Generator Service Guide
.
EEROM Failed !!
: This error will only o ccur if the service
adjustment menu is accessed. Specically, an attempt has b een made
to write to a test patch and EEROM failed to store the data.
ERROR: CALIBRATION FAILED !!
: This error will only o ccur if
the service adjustment menu is accessed. Sp ecically, an A14 sweep
ramp calibration has been attempted and failed. Run the sw
eep
ramp selftest. Refer to the \MENU MAPS" chapter in the
Agilent
Technologies 8360 B-Series Swept Signal Generator/8360 L-Series
Swept CW Generator Service Guide
.
ERROR - Must rst enter correction freq
: This error o ccurs when
a correction p oint does not have its corresp onding frequency entered
rst. Refer to \Creating and Applying the User Flatness Correction
Array" in Chapter 1.
ERROR: Must rst enter a List Frequency !!
: This error o ccurs
when a dwell or oset value does not have its corresp onding
frequency entered rst. Refer to \Creating and Applying the User
Flatness Correction Array" in Chapter 1.
ERROR: Power SearchFailed !!
: This error o ccurs when the ALC
is in the ALC search mo de and is unable to level to the desired
power level. Refer to Chapter 4 and follow the lo cal op erator's check
pro cedures.
ERROR: Start must be
<
Stop !!
: This error o ccurs in asso ciation
with the frequency list, auto ll, feature. If the start frequency
entered is greater than the stop frequency,you will see this error.
Correct byentering a start frequency less than the stop frequency
.
ERROR: Stop must be
>
Start !!
: This error o ccurs in asso ciation
with the frequency list, auto ll, feature. If the stop frequency
2a-2 Error Messages