7 Maintaining the Packed Column Inlet
To Clean the Septum Seat in the Packed Column Inlet
102 Maintaining Your GC
12 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
13 Perform an inlet Leak & Restriction test and reset the maintenance counters.
14 Restore the analytical method.
1 mm