6 Maintaining the Purged Packed Inlet
To Change the Glass Liner on the Purged Packed Inlet
80 Maintaining Your GC
To Change the Glass Liner on the Purged Packed
Inlet
1 Gather the following:
Replacement glass liner, see “Consumables and Parts for the Purged Packed Inlet” on
page 66.
9/16-inch wrench
Lint-free gloves
2 Launch the GC maintenance wizard: Maintenance > Inlets, select the inlet, then select
Perform Maintenance > Replace Liner > Start Maintenance. Wait for the GC to become
ready. The wizard will guide you through the maintenance steps described below.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If the inlet is hot, wear
heat-resistant gloves to protect your hands.
3 Loosen the top insert assembly to remove the top portion of the inlet.
4 Use a thin wire or wood splint to carefully lift and remove the old glass liner.
Top insert weldment
Flared end
Glass liner

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals