6 Maintaining the Purged Packed Inlet
To Change the Septum on the Purged Packed Inlet
74 Maintaining Your GC
To Change the Septum on the Purged Packed Inlet
1 Gather the following:
Replacement septum, see “Consumables and Parts for the Purged Packed Inlet” on
page 66.
Septum nut wrench
0- or 00-grade steel wool (optional)
Tweezers
2 Launch the GC maintenance wizard: Maintenance > Inlets, select the inlet, then select
Perform Maintenance > Replace Septum > Start Maintenance. Wait for the GC to become
ready. The wizard will guide you through the maintenance steps described below.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If the inlet is hot, wear
heat-resistant gloves to protect your hands.
3 Remove the septum retainer nut or Merlin cap.
4 Use tweezers to remove the septum or Merlin Microseal from the retainer nut. Do not
gouge or scratch the interior of the septum head.
5 Firmly press the new septum or Merlin Microseal into the fitting. The metal parts side of
the Merlin Microseal should face down (toward the oven).
Septum retainer nut
Septum

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals