9 Maintaining the FID
To Install the Optional FID PTFE Chimney Insert
160 Maintaining Your GC
To Install the Optional FID PTFE Chimney Insert
WARNING
Be careful! The detector may be hot enough to cause burns. If the detector is hot, wear
heat-resistant gloves to protect your hands.
1 Light the FID flame.
2 Insert the PTFE chimney into the FID castle.
NOTE
When installed, the PTFE chimney insert prevents ignition.