EasyManuals Logo

Agilent Technologies 8860 Maintaining

Agilent Technologies 8860
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #169 background imageLoading...
Page #169 background image
10 Maintaining the TCD
To Bakeout Contaminants from the TCD
Maintaining Your GC 169
To Bakeout Contaminants from the TCD
The bakeout can be performed with the column installed or the detector capped.
CAUTION
If the column is not installed, you must turn off the TCD filament and cap the detector column
fitting to prevent irreparable damage to the filament caused by oxygen entering the detector.
1 If the column is not installed, cap the detector.
WARNING
Be careful! The detector may be hot enough to cause burns. If the detector is hot, wear
heat-resistant gloves to protect your hands.
2 Turn off the TCD filament.
3 If the column is attached to the inlet, maintain inert carrier gas flow through the column.
WARNING
If using hydrogen as a carrier gas, turn off the hydrogen supply and cap the end of the column
to prevent an oven explosion.
4 Set the reference gas flow rate between 20 and 30 mL/min.
5 Set the detector temperature to 375 °C.
6 Hold at 375 °C for several hours.
7 If the column is uninstalled, install it. See “To Install a Capillary Column in the TCD” on
page 166.
8 Load the analytical method.
WARNING
Be careful! The oven or detector fittings may be hot enough to cause burns.
9 Allow the oven, inlet, and detector to equilibrate at operating temperature, then re-tighten
the fittings.

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals