7 Maintaining the Packed Column Inlet
To Install Ferrules on a Packed Metal Column
118 Maintaining Your GC
To Install Ferrules on a Packed Metal Column
1 Gather the following:
Wrenches
Stainless steel male Swagelok fitting, 1/4- or 1/8-inch od
Brass Swagelok nut and ferrule set. See “Consumables and Parts for the Packed
Column Inlet” on page 96.
Lint-free gloves
2 Verify that the column end is cut square and is free of burns and deformation.
3 Secure the fitting in a bench vise.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
4 Assemble a Swagelok nut and ferrules onto the column.
5 Fully insert the column into the vise-held fitting, then withdraw 1–2 mm. Finger-tighten the
nut.
6 Tighten the nut an additional 3/4 turn with a wrench (for a 1/8-inch column) or an
additional 1-1/4 turn (for a 1/4-inch column).
7 Unscrew the column nut from the vise-held fitting and remove the column. Ferrules should
now be set in place on the column with the column end correctly positioned.
Front ferrule
Back ferrule
Column nut

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals