9 Maintaining the FID
To Replace an FID Jet
Maintaining Your GC 147
To Replace an FID Jet
1 Gather the following:
Replacement jet
T-20 Torx screwdriver
1/4-inch nut driver
Tweezers
Compressed, filtered, dry air or nitrogen
Solvent that will clean the type of deposits in your detector
Clean cloth
Cotton swab
Lint-free gloves
2 Launch the GC maintenance wizard: Maintenance > Detectors, select the detector, then
select Perform Maintenance > Replace FID Jet > Start Maintenance. The wizard will guide
you through the maintenance steps described below.
WARNING
Be careful! The oven and/or detector may be hot enough to cause burns. If the detector is
hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
3 If installed, remove the capillary column from the detector.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
4 Remove the FID castle assembly and place it on a clean cloth.
5 Locate the jet inside the housing.

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals