10 Maintaining the TCD
To Install a Capillary Column in the TCD
166 Maintaining Your GC
To Install a Capillary Column in the TCD
1 Gather the following materials, see “Consumables and Parts for the Thermal Conductivity
Detector” on page 164:
Column
Ferrule(s)
Column nut
Column cutter
1/4-inch open-end wrench
Septum
Isopropanol
Lab tissue
Lint-free gloves
2 Launch the GC maintenance wizard: Maintenance > Column > Perform Maintenance >
Install Column > Start Maintenance. Wait for the GC to become ready.
The wizard will guide you through the maintenance steps described below.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
3 Place a septum (if the column id is 0.1 mm), capillary column nut, and ferrule on the
column.
4 Score the column using a glass scribing tool. The score must be square to ensure a clean
break.
Ferrule
Column nut
Septum

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals