EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>8860

Agilent Technologies 8860 Maintaining

Agilent Technologies 8860
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
3 Maintaining the GC
To Change the Hydrogen Sensor Calibration Gas Cylinder
Maintaining Your GC 27
To Change the Hydrogen Sensor Calibration Gas
Cylinder
Replace the calibration gas cylinder when you cannot adjust it to provide the correct flow rate
during calibration, or if it expires.
WARNING
Connect only the Agilent calibration gas to the hydrogen sensor fitting. The calibration gas is
2% hydrogen in air and is not potentially explosive. Gases with higher concentrations of
hydrogen can create a dangerous condition in the oven or damage the hydrogen sensor.
WARNING
Be careful! The oven and/or detector may be hot enough to cause burns. While hydrogen
sensor calibration can be performed at any temperature, cool the oven and detector to safe
handling temperatures, < 40 °C, before continuing.
1 Make sure the old cylinder is empty.
2 Turn the pressure regulator completely off and set the output pressure as low as possible
(full CCW).
3 Remove the pressure regulator from the old cylinder and install it onto the new cylinder.
4 Install the new cylinder into the stand.
5 Turn on the pressure in the new cylinder.
6 Open the GC oven door and connect a flowmeter tube to the sensor tube in the oven.
7 Select Settings > Calibration > Hydrogen Sensor > Start Calibration Cycle? > On/Yes. The
calibration cycle begins. The hydrogen sensor module will wait to stabilize, then will begin
to send calibration gas across the sensor.
8 While continuing to measure the flow rate from the tube, adjust the pressure regulator on
the calibration gas cylinder until the flow rate is approximately 30 mL/min. Remove the
flowmeter and close the oven door.
9 Allow the calibration cycle to complete (approximately 5 minutes total).
10 Check for leaks.
Connect the flow
meter to the hydrogen
sensor tube next to
the oven heater

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals