EasyManuals Logo

Agilent Technologies 8860 Maintaining

Agilent Technologies 8860
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #201 background imageLoading...
Page #201 background image
12 Maintaining the NPD
To Check the NPD Leakage Current
Maintaining Your GC 201
To Check the NPD Leakage Current
1 Load the analytical method.
2 Turn Off the Bead.
Leave the NPD at operating temperature
Leave flows on or off
3 Show the detector output signal in Status.
4 Verify that the output (leakage current) is stable and < 2.0 pA.
The output should slowly drop towards 0.0 pA, and should stabilize in the tenths of a
picoamp. Current > 2.0 pA indicates a problem.

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals