13 Maintaining the FPD+
To Replace the FPD+ Ignitor
Maintaining Your GC 213
To Replace the FPD+ Ignitor
1 Gather the following:
Ignitor replacement kit, see “Consumables and Parts for the FPD+” on page 204.
Torx screwdrivers, T-20 and T-10
5/16-inch nut driver (or wrench)
2 Launch the GC maintenance wizard: Maintenance > Detector > Perform Maintenance >
Replace FPD Ignitor> Start Maintenance. Wait for the GC to become ready.
The wizard will guide you through the maintenance steps described below.
WARNING
Be careful! The detector may be hot enough to cause burns. If the detector is hot, wear
heat-resistant gloves to protect your hands.
3 Remove the FPD cover. See “To Remove the FPD+ Cover” on page 212.
4 Loosen the collet screw holding the cable assembly to the ignitor. Remove the collet and
cable assembly.
5 Use a nut driver to loosen and remove the glow plug.
If using a 5/16-inch wrench, you may need to remove the exit tube assembly using a
T-10 Torx dr iver

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals