5 Maintaining the Split/Splitless Inlet
To Clean the Split/Splitless Inlet
Maintaining Your GC 61
To Clean the Split/Splitless Inlet
1 Gather the following:
Replacement septum, see “Consumables and Parts for the Split/Splitless Inlet” on
page 42.
Replacement liner
Replacement O-ring
Replacement gold seal
Replacement washer
Solvent that will clean the type of deposits in your inlet
Compressed, filtered, dry air or nitrogen
Beaker
Cleaning brushes—The FID cleaning kit (part number 9301-0985) contains appropriate
brushes
Lint-free gloves
2 Manually set the inlet and oven temperature to < 40 °C, and wait for the inlet, oven, and
other parts you might come into contact with inside the oven, to cool before continuing.
Alternately, place the GC in maintenance mode: Maintenance > Instrument > Perform
Maintenance > Maintenance Mode > Start Maintenance.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If the inlet is hot, wear
heat-resistant gloves to protect your hands.
3 Remove the inlet liner, O-ring, and septum. See “To Change the Liner and O-Ring on the
Split/Splitless Inlet” on page 54.
4 Disconnect the column from the inlet.
5 Remove the reducing nut, gold seal, and washer. See “To Replace the Gold Seal on the
Split/Splitless Inlet” on page 57.
6 Place a beaker in the oven under the inlet to catch the solvent.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
7 Soak a cleaning brush in the solvent and scrub the inside of the insert assembly. Repeat
10 times.
8 Rinse the inlet with the solvent.
9 Blow the inside of the inlet dry with compressed air or nitrogen.
10 Install the gold seal, washer, and reducing nut.
11 Install the liner, O-ring, and septum.
12 Install the column. See “To Install a Capillary Column with the Split/Splitless Inlet” on
page 46.