EasyManuals Logo

Agilent Technologies 8860 Maintaining

Agilent Technologies 8860
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #117 background imageLoading...
Page #117 background image
7 Maintaining the Packed Column Inlet
To Condition a Packed Column
Maintaining Your GC 117
To Condition a Packed Column
1 Gather the following:
Capillary adapter, column nut, and no-hole ferrule (for FID), or 1/8-inch Swagelok cap
(for TCD)
Two 7/16-inch wrenches
1/4-inch open-end wrench
Lint-free gloves
WARNING
Do not use hydrogen as the carrier for conditioning! It could vent into the oven and present an
explosion hazard.
2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance. Wait for the GC to become ready.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
3 Install the proper liner in the inlet and attach the column. See “To Install a Packed Metal
Column” on page 111.
4 Cap the detector(s) fittings with the capillary adapter, no-hole ferrule and column nut (FID)
or 1/8-inch cap (TCD).
5 Enter a column flow as recommended by the packing manufacturer or an appropriate flow
as follows:
20 to 30 mL/min for 2-mm id glass or 1/8-inch od metal columns
50 to 60 mL/min for 4-mm id glass or 1/4-inch od metal columns
6 Raise the oven temperature slowly to the conditioning temperature for the column. The
conditioning temperature is never higher than the maximum temperature limit for the
column; 30 °C less than the maximum is usually sufficient.
7 Continue conditioning overnight at the final temperature. Cool the oven to room
temperature with carrier flow on.
8 Attach the column to the detector and maintain established flow. See “To Install a Packed
Metal Column” on page 111.
9 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals