7 Maintaining the Packed Column Inlet
To Clean the Packed Column Inlet
108 Maintaining Your GC
To Clean the Packed Column Inlet
1 Gather the following:
Replacement O-ring. See “Consumables and Parts for the Packed Column Inlet” on
page 96.
Replacement glass liner
Replacement septum
Solvent that will clean the type of deposits in your inlet
Compressed, filtered, dry air or nitrogen
Beaker
Cleaning brushes—The FID cleaning kit (part number 9301-0985) contains appropriate
brushes
Lint-free gloves
2 Manually set the inlet and oven temperature to < 40 °C, and wait for the inlet, oven, and
other parts you might come into contact with inside the oven, to cool before continuing.
Alternately, place the GC in maintenance mode: Maintenance > Instrument > Perform
Maintenance > Maintenance Mode > Start Maintenance.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If the inlet is hot, wear
heat-resistant gloves to protect your hands.
3 Remove the column.
4 Remove the septum nut and septum.
5 Remove the top insert weldment.
6 Remove the glass liner and O-ring.
7 If used, remove the adapter.
8 Ultrasonically clean the septum nut, top insert weldment, and adapter (if used) in a suitable
solvent.
9 Place a beaker in the oven under the inlet to catch the solvent.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
10 Soak a cleaning brush with the solvent and vigorously scrub the interior walls of the inlet.
11 Blow the inside of the inlet dry with compressed air or nitrogen.
12 Install the adapter, if used. See “To Install an Adapter on the Packed Column Inlet” on
page 103.
13 Install the glass liner and O-ring. See “To Change the Glass Liner on the Packed Column
Inlet” on page 105.
14 Install the top insert weldment and finger-tighten.

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals