7 Maintaining the Packed Column Inlet
To Clean the Packed Column Inlet
Maintaining Your GC 109
15 Install the septum and septum nut. See “To Change the Septum on the Packed Column
Inlet” on page 99.
16 Attach the column.
17 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
18 Reset the septum and glass liner counters.
19 Perform an inlet Leak & Restriction test and reset the maintenance counters.
20 Restore the analytical method.