13 Maintaining the FPD+
To Install the FPD+ Cover
Maintaining Your GC 215
To Install the FPD+ Cover
1 Gather the following:
T-20 Torx screwdriver
2 Install the cover.
Single-wavelength detector:
a Start the two screws on the right side of the cover.
b Start and tighten the screws at the base on the left side.
c Tighten the screws on the right side.
Dual-wavelength detector: Install the cover (two screws).
3 Close the FPD detector top cover.