15 Maintaining a Valve
To Remove the Upper Valve Box
236 Maintaining Your GC
To Remove the Upper Valve Box
1 Gather a T-20 Torx screwdriver.
2 Set the valve box to a safe handling temperature (25 °C), or prepare for maintenance:
Maintenance > Instrument > Perform Maintenance > Maintenance Mode > Start
Maintenance. Wait for the GC to become ready.
WARNING
The oven, inlet, detector, and valve box may be very hot. If they are hot, wear heat-resistant
gloves to protect your hands.
3 Lift and remove the detector cover.
4 Remove the mounting screws from the upper valve box.
5 Lift up and set aside.
6 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
WARNING
The valve box insulation is made of refractory ceramic fibers (RCFs). To avoid inhaling RCF
particles, we recommend these safety procedures:
Ventilate your work area
Wear long sleeves, gloves, safety glasses, and a disposable dust/mist respirator
Dispose of insulation in a sealed plastic bag
Vacuum any residual particles and discard
Wash your hands with mild soap and cold water after handling RCFs.

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals