6 Maintaining the Purged Packed Inlet
To Change the Septum on the Purged Packed Inlet
Maintaining Your GC 75
6 Replace the septum retainer nut or Merlin cap and finger-tighten. Tighten the septum
retainer nut until the C-ring is about 1 mm above the nut.
Overtightening the septum nut can cause contamination.
7 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
Restriction tests, and will automatically reset the maintenance counters.
8 Select Finished, then select OK to exit the GC maintenance wizard.
9 Restore the analytical method.
Merlin Microseal and cap
Standard septum