EasyManua.ls Logo

Agilent Technologies 8860 - Page 75

Agilent Technologies 8860
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Maintaining the Purged Packed Inlet
To Change the Septum on the Purged Packed Inlet
Maintaining Your GC 75
6 Replace the septum retainer nut or Merlin cap and finger-tighten. Tighten the septum
retainer nut until the C-ring is about 1 mm above the nut.
CAUTION
Overtightening the septum nut can cause contamination.
7 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
Restriction tests, and will automatically reset the maintenance counters.
8 Select Finished, then select OK to exit the GC maintenance wizard.
9 Restore the analytical method.
Merlin Microseal and cap
Standard septum
1 mm

Table of Contents

Other manuals for Agilent Technologies 8860

Related product manuals