EasyManua.ls Logo

Agilent Technologies 8860 - Page 81

Agilent Technologies 8860
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Maintaining the Purged Packed Inlet
To Change the Glass Liner on the Purged Packed Inlet
Maintaining Your GC 81
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
5 Wearing gloves, inspect the O-ring and replace, if necessary. See “To Change the O-Ring
on the Purged Packed Inlet” on page 79.
6 Wearing gloves, grasp the flared end (top) of the replacement glass liner with tweezers and
install it in the inlet. If the glass liner does not seat properly because a capillary column is
installed, remove the column, install the glass liner, and replace the column. See “To Install
a Capillary Column with the Purged Packed Inlet” on page 70.
7 Install the top insert assembly and hand-tighten firmly.
8 Configure the new liner.
9 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
Restriction tests, and will automatically reset the maintenance counters.
10 Select Finished, then select OK to exit the GC maintenance wizard.
11 Restore the analytical method.

Table of Contents

Other manuals for Agilent Technologies 8860

Related product manuals