EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>8860

Agilent Technologies 8860 Maintaining

Agilent Technologies 8860
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #125 background imageLoading...
Page #125 background image
8 Maintaining the COC Inlet
To Install a Capillary Column with the Cool On-Column Inlet
Maintaining Your GC 125
6 Break off the column end by supporting it against the column cutter opposite the scribe.
Inspect the end with a magnifying loupe to make certain there are no burrs or jagged
edges.
7 Wipe the column walls with a tissue dampened with isopropanol to remove fingerprints
and dust.
8 Gently insert the column into the inlet until it bottoms. You should feel spring tension as
you push up on the column. (Do not withdraw the column.)
9 Insert the column nut into the inlet fitting and finger-tighten.
WARNING
To avoid bending the inlet, always use two wrenches. Use a 5/16-inch wrench to support the
inlet while tightening the column nut with a 1/4-inch wrench.
10 Tighten the column nut an additional 1/4 turn with a wrench or until the column does not
move.
11 If using an automatic injection system with a 0.25-mm or 0.32-mm column, verify that the
column installation by manually pushing the syringe into the inlet.
Good
Bad

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals