8 Maintaining the COC Inlet
To Clean the COC Inlet
Maintaining Your GC 133
8 Remove the spring and insert from the bath and rinse thoroughly with water and methanol.
9 Dry the spring and insert with compressed air or nitrogen.
10 Install the insert. See “To Install an Insert on the COC Inlet” on page 130.
11 Install the column. See “To Install a Capillary Column with the Cool On-Column Inlet” on
page 124.
12 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.