11 Maintaining the ECD
To Bakeout the ECD
Maintaining Your GC 183
8 Check the ECD output value on the control table. It should be lower than the first reading. If
it is not, contact your Agilent service representative.
9 Reinstall the column.
10 Restore the analytical method.