13 Maintaining the FPD+
To Change the FPD+ Wavelength Filter
Maintaining Your GC 211
8 Install the filter in the filter housing. Install the sulfur filter spacer, if necessary.
9 Replace the PMT assembly and secure with the spring.
10 Route the PMT wires through the clips as shown. Avoid placing the wires very near heated
areas (such as the emission block or oven top).
11 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
Restriction tests, and will automatically reset the maintenance counters.
12 Select Finished, then select OK to exit the GC maintenance wizard.
13 Restore the analytical method.