5 Maintaining the Split/Splitless Inlet
To Change the Septum on the Split/Splitless Inlet
Maintaining Your GC 51
CAUTION
Overtightening the septum nut can cause contamination.
7 Select Maintenance > Inlets > Septum injections, then touch Reset Counter.
8 Perform an inlet Leak & Restriction test and reset the maintenance counters.
9 Select Finished, then select OK to exit the GC maintenance wizard.
10 Restore the analytical method.
1 mm