EasyManua.ls Logo

Agilent Technologies 8860 - Operating the Gas Sample Selector

Agilent Technologies 8860
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Maintaining the Split/Splitless Inlet
To Clean the Septum Seat in the Insert Assembly of the Split/Splitless Inlet
Maintaining Your GC 53
6 Scrub the residue from the retainer nut and septum holder with a small piece of rolled-up
steel wool and tweezers. Do not do this over the inlet.
7 Use compressed air or nitrogen to blow away the pieces of steel wool and septum.
8 Line up the tab on the bottom of the insert assembly with the slot on the inlet body and
push down to connect. Slide the locking tab to the left.
9 Firmly press the new septum or Merlin Microseal into the fitting. See “To Change the
Septum on the Split/Splitless Inlet” on page 50.
10 Replace the septum retainer nut or Merlin cap and finger-tighten. See “To Change the
Septum on the Split/Splitless Inlet” on page 50.
11 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
12 Select Maintenance > Inlet > Septum injections, then select Reset.
13 Perform a Leak & Restriction test.
14 Restore the analytical method.

Table of Contents

Other manuals for Agilent Technologies 8860

Related product manuals