5 Maintaining the Split/Splitless Inlet
To Change the Liner and O-Ring on the Split/Splitless Inlet
Maintaining Your GC 55
4 Loosen the O-ring from the sealing surface with tweezers.
5 Grasp the liner with tweezers and pull it out.
6 Inspect the surface of the gold seal for graphite or rubber septum contamination. If
required, replace the gold seal. See“To Replace the Gold Seal on the Split/Splitless
Inlet” on page 57.