EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>8860

Agilent Technologies 8860 Maintaining

Agilent Technologies 8860
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #72 background imageLoading...
Page #72 background image
6 Maintaining the Purged Packed Inlet
To Install a Capillary Column with the Purged Packed Inlet
72 Maintaining Your GC
10 Thread the column nut into the inlet adapter but do not tighten.
11 Adjust the column position so that the septum is even with the bottom of the column nut.
Finger-tighten the column nut until it begins to grip the column.
12 Tighten the column nut an additional 1/4 to 1/2 turn with a wrench so that the column
cannot be pulled from the fitting with gentle pressure.
13 Configure the new column.
14 Condition the column per the manufacturer’s recommendation. See “To Condition a
Capillary Column”.
15 Install the column into the detector.
NOTE
Perform the column installation procedure manually. Do not use the automated column
installation wizard.
To Install a Capillary Column in the FID
To Install a Capillary Column in the NPD
To Install a Capillary Column in the TCD
To Install a Capillary Column in the ECD
To Install a Capillary Column to the FPD+
16 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
Restriction tests, and will automatically reset the maintenance counters.
17 Select Finished, then select OK to exit the GC maintenance wizard.
1%2 mm

Table of Contents

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals