6 Maintaining the Purged Packed Inlet
To Clean the Purged Packed Inlet
Maintaining Your GC 83
NOTE
Perform the installation procedures manually. Do not use the automated installation wizards.
15 Install the septum and septum nut. See “To Change the Septum on the Purged Packed
Inlet” on page 74.
16 Attach the column. See “To Install a Capillary Column with the Purged Packed Inlet” on
page 70.
17 Restore the analytical method.