6 Maintaining the Purged Packed Inlet
To Install a Packed Metal Column
Maintaining Your GC 87
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
21 Allow the oven, inlet, and detector to equilibrate at operating temperature, then retighten
the fittings.
22 Perform the inlet Leak & Restriction test and reset the maintenance counters.

Table of Contents

Other manuals for Agilent Technologies 8860